Similar fixation of total knee arthroplasty components with medial congruent and cruciate retaining polyethylene inserts. A randomised double-blinded controlled radiostereometry trial with 24 months of follow-up

Ortopedia
Imagem de arquivo, ilustrativa — não é do estudo. Radiografia histórica do joelho de um menino (1896). Wellcome Collection · CC BY 4.0 · Wikimedia Commons

Ensaio clínico randomizadoNível 2 · Nova evidênciaKnee surgery, sports traumatology, arthroscopy : official journal of the ESSKA7 jul 20252 citações

Por ScienceMed · · a partir do resumo dos autores

Salvar

Resumo original em inglês

O que aconteceu

To assess the influence of a higher constraint medial congruent (MC) versus a cruciate retaining (CR) (gold standard) polyethylene insert on the migration of cemented tibial and femoral components of the Persona Total Knee System. High-constraint inserts may increase forces on components and potentially affect implant migration and long-term fixation.

A cohort of 66 patients with primary knee osteoarthritis received a cemented Persona total knee arthroplasty and were randomised to either an MC or a CR polyethylene insert. Tibial and femoral component migration was assessed using static radiostereometric analysis, with the first stereo-radiograph taken supine on the first postoperative day (baseline) and again at 3-, 12-, and 24-month follow-up. Tibial and femoral component migration was evaluated in six degrees of freedom and as maximum total point motion (MTPM).

Level I.

O que encontraram

At the 12-month follow-up, the mean difference in tibial component MTPM was 0.04 mm (95% confidence interval [CI]: -0.21 to 0.28). The mean MTPM was 0.81 mm (95% CI: 0.64-0.99) in the MC group and 0.85 mm (95% CI: 0.68-1.03) in the CR group. Signed migrations were similar for the MC and the CR group throughout 24 months of follow-up (p > 0.09). At the 12-month follow-up, the mean difference in femoral component MTPM was 0.18 mm (95% CI: -0.17 to 0.53). The mean MTPM was 1.08 mm (95% CI: 0.83-1.32) in the MC group and 1.26 mm (95% CI: 1.01-1.51) in the CR group. The femoral components with MC insert had 0.18 mm (95% CI: 0.03-0.32) and 0.20 mm (95% CI: 0.01-0.39) more lateral migration at 3- and 12-month follow-up, respectively. At the 24-month follow-up, there was no statistically significant differences in migration for either translations, rotations or MTPM.

Por que isso importa

Both tibial and femoral components had similar and acceptable fixation regardless of the type of polyethylene insert.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Similar fixation of total knee arthroplasty components with medial congruent and cruciate retaining polyethylene inserts. A randomised double-blinded controlled radiostereometry trial with 24 months of follow-up

Carl Christian Holkgaard Burvil, Karina Nørgaard Linde, Maiken Stilling, Torben Bæk Hansen, Jesper Dalsgaard, Søren Rytter, Daan Koppens, Emil Toft Petersen

Descritores MeSH

  • Osteoarthritis, Knee
  • Knee Prosthesis
  • Arthroplasty, Replacement, Knee
  • Follow-Up Studies
  • Humans
  • Knee Joint
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Failure