O que aconteceu

The high failure rates of metal on metal (MoM) large diameter head total hip arthroplasty (LDH THA) and hip resurfacing (HR) prevented their long-term comparisons with regards to clinical outcome. Such knowledge would be important as ceramic LDH bearing is now available. With long-term follow-up, we investigated the difference in 1) patient-reported outcome measures (PROMs); 2) revision and adverse events rates, and 3) metal ion levels between MoM LDH THA and HR.

Forty-eight patients were randomized for LDH THA (24) or HR (24) with the same MoM articulation. At a mean follow-up of 14 years, we compared between groups different PROMs, the number of revisions and adverse events, whole blood Cobalt (Co) and Chromium (Cr) ion levels, and radiographic signs of implant dysfunction.

ClinicalTrials.gov ( NCT04516239 ), August 18, 2020. Retrospectively registered.

O que encontraram

LDH THA (all cases: revised and well-functioning) had significantly better WOMAC (94 versus 85, p = 0.04), and more frequently reported having no limitation (p = 0.04). LDH THA revision rate was 20.8% (5/24) versus 8.3% (2/24) for HR (p = 0.4). Mean Co and Cr ion levels were higher in LDH THA compared to the HR (Co: 3.8 μg/L vs 1.7 μg/L; p = 0.04 and Cr: 1.9 μg/L vs 1.4 μg/L, p = 0.1). On radiographic analyses, 2 LDH THAs showed signs of adverse reaction to metal debris, whereas 1 loose femoral HR component was documented.

Por que isso importa

In the long-term, MoM LDH THA had a high trunnion related revision rate but nonetheless showed better PROMs compared to HR. Provided with a well-functioning modular junction, non-MoM LDH THA would offer an appealing option.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Despite higher revision rate, MoM large-head THA offers better clinical scores than HR: 14-year results from a randomized controlled trial involving 48 patients

Lazaros Kostretzis, Martin Lavigne, Marc-Olivier Kiss, Maged Shahin, Janie Barry, Pascal-André Vendittoli

Descritores MeSH

  • Hip Prosthesis
  • Arthroplasty, Replacement, Hip
  • Metal-on-Metal Joint Prostheses
  • Chromium
  • Reoperation
  • Prosthesis Design
  • Cobalt
  • Humans

Veja também

Estudos relacionados