O que aconteceu

Cardiac output monitoring is recommended for high-risk surgical patients and critically ill patients with circulatory shock. We performed a systematic review and meta-analysis of clinical studies published since 2010 that compared minimally invasive pulse wave analysis-derived cardiac output or cardiac index measurements with reference measurements by pulmonary artery thermodilution or transpulmonary thermodilution in adult surgical or critically ill patients.

In a random-effects meta-analysis, we calculated pooled estimates of the percentage error, mean difference and standard deviation, and 95% limits of agreement separately for studies reporting cardiac output or cardiac index. Subgroup analyses were performed by patient population and test device.

PROSPERO (CRD420251090806; submitted July 17, 2025).

O que encontraram

We included 92 studies divided into 113 data sets with a total of 3111 patients. For 71 data sets reporting cardiac output, the pooled percentage error (95% confidence interval [95% CI]) was 44.0% (38.2%-49.8%) with a mean difference (standard deviation) of -0.1 (1.3) L min-1 with 95% limits of agreement of -2.6 to 2.4 L min-1 (I2=9.8%). For 42 data sets reporting cardiac index, the pooled percentage error was 49.1% (40.6%-57.6%) with a mean difference of -0.1 (0.9) L min-1 m-2 with 95% limits of agreement of -1.8 to 1.6 L min-1 m-2 (I2=7.9%). The percentage error varied substantially across patient populations and devices. Overall risk of bias was low.

Por que isso importa

The pooled percentage error between minimally invasive pulse wave analysis-derived cardiac output measurements of 44.0% (cardiac output) and 49.1% (cardiac index) in adult surgical or critically ill patients exceeds the 30% threshold for clinically acceptable agreement. However, the percentage error varied depending on the patient population and device used.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Agreement of minimally invasive pulse wave analysis with pulmonary artery and transpulmonary thermodilution cardiac output measurements in perioperative and intensive care medicine: a systematic review and meta-analysis

Moritz Flick, Dominik X Müller, Alina Bergholz, Pawel Sierzputowski, Michael Eichlseder, Rami Saad, Gillis Greiwe, Karim Kouz, Alexander Hapfelmeier, Bernd Saugel

Descritores MeSH

  • Pulse Wave Analysis
  • Critical Care
  • Pulmonary Artery
  • Perioperative Care
  • Cardiac Output
  • Thermodilution
  • Humans
  • Monitoring, Physiologic
  • Critical Illness

Veja também

Estudos relacionados