Safety profiles of bone-conduction hearing implants revisited: A meta-analytic comparison adjusted for follow-up time
NeurologiaResumo original em inglês
O que aconteceu
To estimate incidence rates of adverse events associated with bone-conduction hearing implants from primary literature and to compare rates among different technological designs.
A systematic literature review and meta-regression was conducted to estimate incidence rates of minor and major complications and their consequences (i.e., revision surgery, explantation, re-implantation and becoming a non-user) while testing for effects of device design, age group, mean follow-up time and study type. These four designs of bone-conduction systems were included: 1) active transcutaneous with electromagnetic transducer (aBCIem), 2) active transcutaneous with piezoelectric transducer (aBCIpz), 3) passive transcutaneous (tBAHA), and 4) passive percutaneous (pBAHA).
O que encontraram
The final dataset included 170 articles reporting on 6451 implantations and 1847 minor and 668 major events. Mean follow-up time was a significant predictor of incidence rates (p < 0.001), with lower rates reported in studies with longer follow-up times. After adjusting to the median follow-up time, the pooled incidence rate of minor complications was significantly lower in aBCIem (p < 0.05) compared to other designs. For both major events and revision surgery, pooled incidence rates were significantly higher in pBAHA compared to aBCIem (p < 0.001) and tBAHA (p < 0.001), but not compared to aBCIpz (Major: p = 0.197; Revision: p = 0.248). Becoming a non-user occurred significantly more frequently in tBAHA compared to other designs (p < 0.005). No statistically significant differences were found in rates of explantation and explantation with re-implantation.
Por que isso importa
When comparing across multiple studies, adverse event rates should be adjusted for different lengths of follow-up. Synthesizing published evidence without considering follow-up time may lead to false conclusions.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Safety profiles of bone-conduction hearing implants revisited: A meta-analytic comparison adjusted for follow-up time
Marco Caversaccio, Wilhelm Wimmer, Annegret Hoch, Thomas Dejaco, Burkard Schwab
Descritores MeSH
- Postoperative Complications
- Hearing Aids
- Bone Conduction
- Reoperation
- Humans
- Follow-Up Studies