To compare the accuracy, efficiency, and safety of robot-assisted versus frame-based stereoelectroencephalography (sEEG) in patients with drug-resistant epilepsy.
In accordance with the PRISMA guidelines, a comprehensive literature search was conducted across four databases (PubMed, Scopus, Web of Science, and Cochrane) up to September 2025. We included comparative studies that evaluated robot-assisted versus frame-based sEEG in patients with drug-resistant epilepsy. A random-effects model was applied to calculate the mean difference (MD) and risk ratio (RR) with corresponding 95% confidence intervals (CI).
O que encontraram
Eight retrospective comparative cohort studies (758 patients) were included. Regarding accuracy, there was no significant difference between the robot-assisted and frame-based sEEG in depth error (MD: 0.24 mm; 95% CI: -0.79 to 1.27), radial error (MD: 0.07 mm; 95% CI: -0.70 to 0.84), entry point error (EPE; MD: -1.35 mm; 95% CI: -2.74 to 0.04), and target point error (MD: -0.02 mm; 95% CI: -0.57 to 0.53). Robot-assisted sEEG demonstrated a significantly shorter overall operation time (MD: -32.58 min; 95% CI: -47.92 to -17.24) and operation time per electrode (MD: -6.55 min; 95% CI: -8.08 to -5.02). However, pre-implantation time (MD: -1.46 min; 95% CI: -14.02 to 11.11) and electrode number per patient (MD: 0.86; 95% CI: -0.84 to 2.56) were comparable between groups. There was no significant difference between the two groups in overall complication rates, including hemorrhagic events, neurological deficits, infections, and technical complications.
Por que isso importa
Robot-assisted sEEG significantly reduced both overall operation time and operation time per electrode compared with the frame-based group. Both techniques demonstrated comparable accuracy and safety profiles.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Robot-assisted versus frame-based stereoelectroencephalography (sEEG) electrode implantation in drug-resistant epilepsy: a meta-analysis of accuracy, efficiency, and safety
Abdallah Abbas, Haneen Sabet, Karima El Refaei, Abrar AbuHamdia, Toka Elboraay, Yasmin Negida, Majed Aldehri, Ibrahim Alnaami, Ahmed M Raslan
Descritores MeSH
- Drug Resistant Epilepsy
- Electrodes, Implanted
- Electroencephalography
- Stereotaxic Techniques
- Robotic Surgical Procedures
- Humans
Veja também
Estudos relacionados
- Usefulness of Robotic Stereotactic Assistance (ROSA®) Device for Stereoelectroencephalography Electrode Implantation: A Systematic Review and Meta-analysis
- Comparative assessment of stereoelectroencephalography and subdural electrodes in invasive epilepsy monitoring: a systematic review and meta-analysis
- Stereoelectroencephalography versus subdural electrodes for invasive monitoring of drug-resistant epilepsy patients: a systematic review and meta-analysis
- Robot-assisted vs. manually guided stereoelectroencephalography for refractory epilepsy: a systematic review and meta-analysis