Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control
PneumologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
gov under the registration no. NCT03785899.
• Previous studies examined SpO2 signal loss (SL) during routine manual oxygen control being rare, but pronounced in lower SpO2 states. • Oxygen titration during SL is unlikely to be beneficial as SpO2 may recover to a normoxic range.
O que encontraram
• Periods of low signal quality of SpO2 are common during pHEs and while supported with automated oxygen control (SPOC), FiO2 is set to a back-up value reflecting FiO2 requirements during normoxemia in response to SL, although SpO2 remained below target until signal recovery. • FiO2 overshoots following pHEs were rare during AOC and occurred with a delayed onset; therefore, increased FiO2 during SL does not necessarily lead to overshoots.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control
Lukas O Langanky, Karen B Kreutzer, Christian F Poets, Axel R Franz, Christoph E Schwarz
Descritores MeSH
- Cross-Over Studies
- Hypoxia
- Infant, Premature
- Oximetry
- Oxygen Inhalation Therapy
- Oxygen Saturation
- Oxygen
- Infant, Premature, Diseases
- Algorithms
- Male
Veja também
Estudos relacionados
- Oxygen Saturation Targeting for Infants with Bronchopulmonary Dysplasia: A Pilot Randomized Trial
- Comparison of two different oxygen saturation target ranges for automated oxygen control in preterm infants: a randomised cross-over trial
- Control of breathing in preterm infants on incubator oxygen or nasal cannula oxygen
- Performance and safety of the PRICO closed-loop oxygen saturation targeting system in neonates: pragmatic multicentre cross-over study (TarOx Study)