Predictive value of the PRAETORIAN score for defibrillation test success in patients with subcutaneous ICD: A subanalysis of the PRAETORIAN-DFT trial
CardiologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
The PRAETORIAN score estimates the risk of failure of subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy by using generator and lead positioning on bidirectional chest radiographs. The PRospective randomized compArative trial of subcutanEous implanTable cardiOverter-defibrillatoR ImplANtation with and without DeFibrillation Testing (PRAETORIAN-DFT) investigates whether PRAETORIAN score calculation is noninferior to defibrillation testing (DFT) with regard to first shock efficacy in spontaneous events.
This prespecified subanalysis assessed the predictive value of the PRAETORIAN score for defibrillation success in induced ventricular arrhythmias.
This multicenter investigator-initiated trial randomized 965 patients between DFT and PRAETORIAN score calculation after de novo S-ICD implantation. Successful DFT was defined as conversion of induced ventricular arrhythmia in <5 seconds from shock delivery within 2 attempts. Bidirectional chest radiographs were obtained after implantation. The predictive value of the PRAETORIAN score for DFT success was calculated for patients in the DFT arm.
O que encontraram
In total, 482 patients were randomized to undergo DFT. Of these patients, 457 (95%) underwent DFT according to protocol, of whom 445 (97%) had successful DFT and 12 (3%) had failed DFT. A PRAETORIAN score of ≥90 had a positive predictive value of 25% for failed DFT, and a PRAETORIAN score of <90 had a negative predictive value of 99% for successful DFT. A PRAETORIAN score of ≥90 was the strongest independent predictor for failed DFT (odds ratio 33.77; confidence interval 6.13-279.95; P < .001).
Por que isso importa
A PRAETORIAN score of <90 serves as a reliable indicator for DFT success in patients with S-ICD, and a PRAETORIAN score of ≥90 is a strong predictor for DFT failure.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Predictive value of the PRAETORIAN score for defibrillation test success in patients with subcutaneous ICD: A subanalysis of the PRAETORIAN-DFT trial
Reinoud E Knops, Mikhael F El-Chami, Christelle Marquie, Peter Nordbeck, Anne-Floor B E Quast, Roland R Tilz, Tom F Brouwer, Pier D Lambiase, Christopher J Cassidy, Lucas V A Boersma et al.
Descritores MeSH
- Defibrillators, Implantable
- Predictive Value of Tests
- Electric Countershock
- Male
- Middle Aged
- Prospective Studies
- Ventricular Fibrillation
- Death, Sudden, Cardiac
- Tachycardia, Ventricular
- Aged
Veja também
Estudos relacionados
- Catheter Ablation for Ventricular Tachycardia in Ischemic Cardiomyopathy: A Meta-Analysis With Reconstructed Time-to-Event and Trial Sequential Analysis
- Incidence and predictors of major arrhythmic events after acute myocarditis complicated by ventricular arrhythmias: A systematic review and meta-analysis
- Time to Benefit of Epicardial Ablation in Malignant Forms of Brugada Syndrome: A Reconstructed Individual Patient Data Meta-Analysis
- Wearable cardioverter defibrillator for transient arrhythmic risk and sudden cardiac death prevention: a systematic review and updated meta-analysis
- The Predictive Value of Impedance for Defibrillation Success After S-ICD Implantation: A Reliable Indicator?