New possibilities for ST analysis - A post-hoc analysis on the Dutch STAN RCT
EndocrinologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
The diagnostic value of ST analysis of the fetal electrocardiogram (fECG) during labor is uncertain. False alarms (ST events) may be explained by physiological variation of the fetal electrical heart axis. Adjusted ST events, based on a relative rather than an absolute rise from baseline, correct for this variation and may improve the diagnostic accuracy of ST analysis.
Determine the optimal cut-off for relative ST events in fECG to detect fetal metabolic acidosis.
Post-hoc analysis on fECG tracings from the Dutch STAN trial (STAN+CTG branch).
1328 term singleton fetuses with scalp ECG tracing during labor, including 10 cases of metabolic acidosis.
O que encontraram
Cut-off value for relative ST events at the point closest to (0,1) in the receiver operating characteristic (ROC) curve with corresponding sensitivity and specificity.
Relative baseline ST events had an optimal cut-off at an increment of 85% from baseline. Relative ST events had a sensitivity of 90% and specificity of 80%.
Por que isso importa
Adjusting the current definition of ST events may improve ST analysis, making it independent of CTG interpretation.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
New possibilities for ST analysis - A post-hoc analysis on the Dutch STAN RCT
A D J Hulsenboom, M B Van der Hout-van der Jagt, E S A van den Akker, P C A M Bakker, E van Beek, A P Drogtrop, A Kwee, M E M H Westerhuis, R J P Rijnders, N W E Schuitemaker et al.
Descritores MeSH
- Labor, Obstetric
- Acidosis
- Female
- Fetal Heart
- Fetal Monitoring
- Heart Rate, Fetal
- Humans
- Pregnancy
- Cardiotocography
- Electrocardiography
Veja também
Estudos relacionados
- Fetal Heart Rate Auscultation, 4th Edition
- ST waveform analysis vs cardiotocography alone for intrapartum fetal monitoring: An updated systematic review and meta-analysis of randomized trials
- Computerized analysis of cardiotocograms in clinical practice and the SisPorto® system thirty-two years after: technological, physiopathological and clinical studies
- Comparison of effect of CTG + STan with CTG alone on emergency Cesarean section rate: STan Australian Randomized controlled Trial (START)