Left ventricular electrical delay predicts volumetric response to leadless cardiac resynchronization therapy

Cardiologia
Imagem de arquivo, ilustrativa — não é do estudo. Músculo cardíaco em corte transversal, ao microscópio óptico. Berkshire Community College Bioscience Image Library · CC0 · Wikimedia Commons

Ensaio clínico randomizadoNível 2 · Nova evidênciaHeart rhythm28 ago 20242 citações

Por ScienceMed · · a partir do resumo dos autores

Salvar

Resumo original em inglês

O que aconteceu

Leadless left ventricular (LV) endocardial pacing is an emerging cardiac resynchronization therapy (CRT) technology. Predictors of response to leadless CRT are poorly understood. Implanting the LV endocardial pacing electrode in sites with increased electrical latency (Q-LV) may improve response rates.

The purpose of this study was to examine the association between Q-LV and echocardiographic remodeling response to leadless CRT delivered with the WiSE-CRT system.

A post hoc analysis (n = 122) of the SOLVE-CRT trial examined the relationship between LV pacing site Q-LV with rate of left ventricular end-systolic volume (LVESV) reduction >15% at 6 months. Multivariable regression analysis, adjusting for age, sex, previous CRT nonresponse, cardiomyopathy etiology, QRS morphology, and QRS duration was performed, followed by receiver operating characteristic analysis and analysis of variance by Q-LV quartile. A subgroup analysis of the ischemic cardiomyopathy cohort was undertaken.

O que encontraram

Complete Q-LV data were available for 122 of 153 patients (80%) in the active arms SOLVE-CRT. Overall, the 6-month LVESV response rate was 46%. Logistic regression identified Q-LV as an independent response predictor with borderline significance (adjusted odds ratio 1.015; P = .05). Analysis by Q-LV quartile demonstrated a significant improvement in response rate in quartile 4 (longest Q-LV 64%) compared to quartile 1 (shortest Q-LV 28%) (P <.01). This association was primarily driven by strong Q-LV-response correlation in patients with ischemic cardiomyopathy, demonstrated by subgroup logistic regression (adjusted odds ratio 1.034; P = .004).

Por que isso importa

Increased Q-LV was associated with improved reverse remodeling following leadless CRT. Targeting LV endocardial sites of high Q-LV may deliver additional benefit compared to empirical LV electrode implantation.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Left ventricular electrical delay predicts volumetric response to leadless cardiac resynchronization therapy

Nadeev Wijesuriya, Vishal Mehta, Felicity De Vere, Sandra Howell, Nilanka Mannakkara, Baldeep Sidhu, Mark Elliott, Paolo Bosco, Prashanthan Sanders, Jagmeet P Singh et al.

Descritores MeSH

  • Heart Failure
  • Stroke Volume
  • Heart Ventricles
  • Ventricular Remodeling
  • Cardiac Resynchronization Therapy
  • Aged
  • Male
  • Middle Aged
  • Treatment Outcome
  • Humans

Veja também

Estudos relacionados