ScienceMed®

Incidence and risk factors for central venous access device failure in hospitalized adults: A multivariable analysis of 1892 catheters

Infectologia

Ensaio clínico randomizadoNível 2 · Nova evidênciaJournal of hospital medicine27 mai 202420 citações

Por ScienceMed · · a partir do resumo dos autores

Salvar

Resumo original em inglês

O que aconteceu

Central venous access devices (CVADs) allow intravenous therapy, haemodynamic monitoring and blood sampling but many fail before therapy completion.

To quantify CVAD failure and complications; and identify risk factors.

Secondary analysis of multicentre randomised controlled trial including patients aged ≥16 years with a non-tunnelled CVAD (NTCVAD), peripherally-inserted central catheter (PICC) or tunnelled CVAD (TCVAD). Primary outcome was incidence of all-cause CVAD failure (central line-associated bloodstream infection [CLABSI], occlusion, accidental dislodgement, catheter fracture, thrombosis, pain). Secondary outcomes were CLABSI, occlusion and dislodgement. Cox regression was used to report time-to-event associations.

O que encontraram

In 1892 CVADs, all-cause failure occurred in 10.2% of devices: 49 NTCVADs (6.1%); 100 PICCs (13.2%); 44 TCVADs (13.4%). Failure rates for CLABSI, occlusion and dislodgement were 5.3%, 1.8%, and 1.7%, respectively. Independent CLABSI predictors were blood product administration through PICCs (hazard ratio (HR) 2.62, 95% confidence interval (CI) 1.24-5.55); and in TCVADs, one or two lumens, compared with three to four (HR 3.36, 95%CI 1.68-6.71), intravenous chemotherapy (HR 2.96, 95%CI 1.31-6.68), and diabetes (HR 3.25, 95%CI 1.40-7.57). Independent factors protective for CLABSI include antimicrobial NTCVADs (HR 0.23, 95%CI 0.08-0.63) and lipids in TCVADs (HR 0.32, 95%CI 0.14-0.72). NTCVADs inserted at another hospital (HR 7.06, 95%CI 1.48-33.7) and baseline infection in patients with PICCs (HR 2.72, 95%CI 1.08-6.83) were predictors for dislodgement. No independent occlusion predictors were found. Modifiable risk factors were identified for CVAD failure, which occurred for 1-in-10 catheters. Strict infection prevention measures and improved CVAD securement could reduce CLABSI and dislodgement risk.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Incidence and risk factors for central venous access device failure in hospitalized adults: A multivariable analysis of 1892 catheters

Amanda Corley, Ruth H Royle, Nicole Marsh, Emily Larsen, E Geoffrey Playford, Matthew R McGrail, Naomi Runnegar, Robert S Ware, Nicole C Gavin, Evan Alexandrou et al.

Descritores MeSH

  • Equipment Failure
  • Catheterization, Central Venous
  • Catheter-Related Infections
  • Central Venous Catheters
  • Adult
  • Male
  • Middle Aged
  • Risk Factors
  • Incidence
  • Humans

Veja também

Estudos relacionados