In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating
OrtopediaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
Fractures of the proximal humerus are common, particularly in elderly populations. Anatomical locking plates target stabilization with a multitude of screws spanning into the humeral head. Sound implant placement and screw length determination are key for a successful clinical outcome but are difficult to obtain from planar X-rays. A novel implant positioning technology for proximal humerus plating (Xin1) outputs screw lengths suggestions and plate position based on hole projections in conventional X-ray images. This study investigated the performance of a prototype Xin1 system in a postmortem (in vitro) experiment as well as in a clinical handling test.
For in vitro testing, twelve shoulders from six anatomical specimens were randomized into two groups to compare the Xin1 technique to the conventional operation in terms of surgical precision, procedure time and X-ray exposure. For the clinical trial, 11 patients undergoing plating of the proximal humerus were included. The aim was to investigate clinical handling of the Xin1 marker clip and to retrospectively evaluate the system performance in a real-life fracture situation. Image pairs before and after insertion of the proximal screws were retrospectively processed to investigate the influence of potential bone fragment shifts on the system output.
O que encontraram
In the postmortem experiment, the use of the system significantly improved the surgical precision (52% error reduction), procedure time (38% shorter) and radiation exposure (64% less X-rays). Clinical handling demonstrated seamless embedding of the marker clip into existing clinical workflows without adverse events reported. Retrospective X-ray analysis on six eligible patients revealed differences in the calculated screw lengths of ≤2 mm before and after screw insertion for five patients. In one patient, the screw lengths differed up to 8 mm, which might indicate displacement of the head fragment.
Por que isso importa
Results suggest a strong potential of the Xin1 assistance technology to enhance the surgical procedure and patient outcomes in the rising incidence of osteoporotic humeral fractures. Robust performance in a real-life fracture situation was observed. In-depth validation of the system is, however, needed before placing it into clinical practice.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating
Markus Windolf, Dominik Knierzinger, Stefaan Nijs, An Sermon, Michael Blauth, Robert Geoff Richards, Jan Buschbaum
Descritores MeSH
- Shoulder Fractures
- Biomechanical Phenomena
- Bone Plates
- Fracture Fixation, Internal
- Humans
- Humerus
- Retrospective Studies
- Shoulder
- Aged
- In Vitro Techniques
Veja também
Estudos relacionados
- Role of Fibular Allograft in Proximal Humerus Fractures: A Systematic Review
- The forgotten fragment: additional lesser tuberosity fixation of 4-part proximal humeral fractures-a biomechanical investigation
- Fracture dislocations of the proximal humerus treated with open reduction and internal fixation: a systematic review
- Improved outcomes for proximal humerus fracture open reduction internal fixation augmented with a fibular allograft in elderly patients: a systematic review and meta-analysis