Glucose Testing Methods: A Systematic Review and Meta-Analysis of Diagnostic Accuracy of Point-of-Care Devices for Neonatal Hypoglycemia
EndocrinologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
To evaluate the accuracy of various point-of-care device methodologies for measuring blood glucose concentrations in babies at risk of neonatal hypoglycemia.
This systematic review and meta-analysis included studies from Ovid MEDLINE, Ovid Embase, and Web of Science up to May 20, 2024. Studies comparing point-of-care testing methods for neonatal blood glucose to a standard laboratory method were included, excluding those on continuous glucose monitoring or conducted before 1990. Two researchers independently assessed inclusion and evaluated risk of bias using QUADAS-2. Sensitivity and specificity were calculated using contingency tables, and diagnostic accuracy was analyzed using hierarchical random-effects modelling. Studies with insufficient data were summarized by estimation direction.
PROSPERO on December 29, 2023 (CRD42023488539).
O que encontraram
Seventy-one studies were included. The quantitative analysis (n = 31) evaluated glucose oxidase (GO) + photometry (n = 8), glucose-1-dehydrogenase (GDH) + photometry (n = 6), GO + electrochemistry (n = 13), GDH + electrochemistry (n = 12), and hexokinase (HK) + electrochemistry (n = 2). All methods showed high specificity (≥93%), with GO + electrochemistry, GDH + electrochemistry, and HK + electrochemistry showing superior sensitivity. The summary receiver operating characteristic curve confirmed HK + electrochemistry as the most accurate.
Por que isso importa
Certain point-of-care device methodologies demonstrate greater accuracy in measuring neonatal blood glucose concentrations. Of the methods evaluated, HK + electrochemistry proved to be the most reliable. However, the limited number of studies using this method suggests the need for further research to confirm these findings across diverse settings and populations.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Glucose Testing Methods: A Systematic Review and Meta-Analysis of Diagnostic Accuracy of Point-of-Care Devices for Neonatal Hypoglycemia
Sophie L St Clair, Caitlyn M Ulyatt, Maria T Corkin, Libby G Lord, Caroline A Crowther, Jane E Harding, Luling Lin
Descritores MeSH
- Point-of-Care Testing
- Blood Glucose
- Hypoglycemia
- Infant, Newborn, Diseases
- Point-of-Care Systems
- Humans
- Sensitivity and Specificity
- Infant, Newborn
Veja também
Estudos relacionados
- Glucômetro versus método enzimático padrão para diabetes mellitus gestacional: revisão sistemática para acurácia do teste diagnóstico
- Diagnostic accuracy of point-of-care ultrasound in neonatal intensive care units: a systematic review
- Effects of Continuous Versus Intermittent Glucose Monitoring in Intensive Care Unit Patients: A Systematic Review With Meta-Analysis
- Point-of-care Tests for Chlamydia trachomatis and Neisseria gonorrhoeae: Review of the Literature