Resumo original em inglês

O que aconteceu

The MANTA is a dedicated plug-based large-bore vascular closure device (VCD) providing safe hemostasis in most patients, but data on the clinical impact and mechanisms of MANTA related complications are limited. This study sought to determine the frequency, impact and predictors of MANTA-related access complications.

This patient-level meta-analysis included data from 2 medical device approval studies and 1 post-approval registry. The primary endpoint was the composite of major and minor access complications. Technical success was defined as hemostasis with MANTA closure device without need for vascular surgery or stenting.

O que encontraram

Eight hundred ninety-one patients (mean age 80) underwent transcatheter aortic valve replacement (n = 814), endovascular aortic repair (n = 71), balloon aortic valvuloplasty (n = 4) or mechanical circulatory support (n = 2). Technical success was 96.4% and median time to hemostasis was 31 (interquartile range: 17-76) seconds. The primary endpoint occurred 9.1% and bailout vascular surgery or stenting was necessary in 32 patients (3.6%). Female gender (OR: 2.63, CI: 1.46-4.73, p = 0.001), left femoral access (OR: 2.18, CI: 1.17-4.06, p = 0.015) and unfavorable arteriotomy phenotype (combination of a small femoral artery diameter with a deep arteriotomy; OR 2.27: 1.26-4.10, p = 0.006) independently predicted access complications. Access complications most often consisted of vessel dissection, stenosis or occlusion and predominantly occurred in patients with an unfavorable arteriotomy phenotype.

Por que isso importa

Large-bore arteriotomy closure with MANTA VCD provided fast and safe hemostasis with an acceptable complication rate. Refined procedure planning and risk-stratification may further improve MANTA VCD performance.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Frequency, Impact, and Predictors of Access Complications With Plug-Based Large-Bore Arteriotomy Closure - A Patient-Level Meta-Analysis

Rutger-Jan Nuis, David Wood, Herbert Kroon, Maarten van Wiechen, Darra Bigelow, Chris Buller, Joost Daemen, Peter de Jaegere, Zvonimir Krajcer, John Webb et al.

Descritores MeSH

  • Aortic Valve Stenosis
  • Catheterization, Peripheral
  • Endovascular Procedures
  • Vascular Closure Devices
  • Hemorrhage
  • Hemostatic Techniques
  • Humans
  • Treatment Outcome
  • Female
  • Femoral Artery

Veja também

Estudos relacionados