Eye Lens Radiation Exposure During TAVI: Current Evidence and Imaging-Based Strategies for Dose Reduction

Radiologia e Imagem
Imagem de arquivo, ilustrativa — não é do estudo. Artéria humana em corte transversal, ao microscópio óptico. Lord of Konrad · CC0 · Wikimedia Commons

Revisão sistemáticaNível 1 · Evidência consolidadaTomography (Ann Arbor, Mich.)4 mar 20260 citações

Por ScienceMed · · a partir do resumo dos autores

Salvar

Resumo original em inglês

O que aconteceu

Transcatheter aortic valve implantation (TAVI) is increasingly performed in fluoroscopy-intensive environments, raising concerns about occupational eye lens dose (equivalent dose to the eye lens, Hp (3)) and the risk of radiation-induced cataract, particularly after the reduction of recommended annual eye lens dose limits to 20 mSv.

To summarize evidence on eye lens radiation exposure during TAVI, identify procedural and occupational determinants, and review strategies to reduce exposure with a focus on imaging optimization.

We performed a narrative review of observational and prospective studies reporting direct eye-level dose measurements or validated surrogate eye lens dose estimates (head-level, chest-level, or DAP-normalized) during TAVI and related structural heart procedures. This approach was chosen to provide a qualitative synthesis of the available evidence rather than a formal systematic review.

O que encontraram

Reported operator eye lens doses typically ranged from 30 to 110 µSv per procedure, with higher exposure during transapical/transaortal access and among staff working close to the patient (e.g., anesthesiologists and circulating nurses). Additional shielding and lead-free drapes reduced normalized eye dose by approximately 25-40%, and RADPAD® use reduced operator eye-level dose from 24.3 to 14.8 µSv per procedure (p = 0.008). At these levels, cumulative exposure may approach recommended regulatory limits after approximately 150-300 procedures, depending on role, access route, and shielding practices.

Por que isso importa

In conclusion, Occupational eye lens exposure during TAVI is clinically relevant and strongly influenced by access route, staff positioning, and imaging-system use. Dose reduction should combine routine eye protection and dedicated eye-level dosimetry with imaging optimization (low pulse-rate fluoroscopy, minimized Digital-Subtraction-Angiography (DSA)/cine acquisitions, tight collimation, avoidance of unnecessary magnification, and correct positioning of ceiling-suspended shields and table skirts).

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Eye Lens Radiation Exposure During TAVI: Current Evidence and Imaging-Based Strategies for Dose Reduction

Chiara Zanon, Alessandro Fiocco, Vincenzo Tarzia, Emilio Quaia

Descritores MeSH

  • Radiation Exposure
  • Radiography, Interventional
  • Occupational Exposure
  • Transcatheter Aortic Valve Replacement
  • Lens, Crystalline
  • Radiation Protection
  • Fluoroscopy
  • Humans
  • Radiation Dosage

Veja também

Estudos relacionados