Evaluation of a research hearing aid for audiological testing

Otorrinolaringologia
Imagem de arquivo, ilustrativa — não é do estudo. Radiografia de tórax de pulmões saudáveis. N.a. Radiology by Dr. Johan Verbeken. Scan and image editing by Vincent Mia Edie Verheyen · domínio público · Wikimedia Commons

Ensaio clínico randomizadoNível 2 · Nova evidênciaInternational journal of audiology26 nov 20243 citações

Por ScienceMed · · a partir do resumo dos autores

Salvar

Resumo original em inglês

O que aconteceu

Open-source hearing aid (HA) research tools provide avenues for testing new audiological concepts. This study compared a wearable research HA (RHA) - the "Portable Hearing Laboratory" - to a high-end commercial HA (CHA) in terms of aided outcome. Using this RHA, it also evaluated a fitting approach based on non-individualised gain presets.

Using a randomised repeated-measures design, participants completed speech recognition and speech quality assessments under unaided and several aided (RHA and CHA with omnidirectional and directional microphone settings) conditions. Furthermore, two approaches for prescribing amplification were evaluated with the RHA: (1) a preset approach based on non-verified clinical gain targets calculated for three standard audiograms, and (2) a best-practice approach based on individually prescribed and verified clinical gain targets.

Fifteen adults aged 50-78 years with symmetrical, mild-to-moderate sensorineural hearing losses.

O que encontraram

The RHA and CHA gave similar audibility, speech recognition in quiet and in noise, and speech quality ratings in quiet and in noise. Relative to the best-practice approach, the preset approach resulted in less audibility but similar speech recognition in quiet and in noise.

Por que isso importa

The RHA appears suited for testing new audiological concepts such as preset-based approaches to HA fitting.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Evaluation of a research hearing aid for audiological testing

Pernilla K Andersen, Lukas Jürgensen, Florian Denk, Hendrik Husstedt, Tim Jürgens, Tobias Neher

Descritores MeSH

  • Wearable Electronic Devices
  • Hearing Aids
  • Hearing Loss, Sensorineural
  • Correction of Hearing Impairment
  • Speech Perception
  • Persons with Hearing Disabilities
  • Perceptual Masking
  • Humans
  • Male
  • Middle Aged

Veja também

Estudos relacionados