Diagnostic performance of qualitative and quantitative methods of meibomian gland dropout evaluation in dry eye disease: An investigator-masked, randomised crossover study

Oftalmologia
Imagem de arquivo, ilustrativa — não é do estudo. As camadas de células nervosas da retina, em microscopia de fluorescência. NIH · domínio público · Wikimedia Commons

Ensaio clínico randomizadoNível 2 · Nova evidênciaContact lens & anterior eye : the journal of the British Contact Lens Association17 nov 20248 citações

Por ScienceMed · · a partir do resumo dos autores

Salvar

Resumo original em inglês

O que aconteceu

To evaluate the discriminative abilities and optimal cut-off values of qualitative meiboscale grading and percentage meibomian gland dropout measurements from each of two instruments (Keratograph 5M and LipiView II) in detecting dry eye disease, as defined by the TFOS DEWS II criteria.

A total of 227 community residents (143 females, 84 males; mean ± SD age, 36 ± 15 years) were recruited in a prospective, investigator-masked, randomised, crossover study. Dry eye symptomology, tear film parameters, and ocular surface characteristics were evaluated in a single clinical session. Qualitative meiboscale grading and quantitative percentage gland dropout assessment using Image J software were evaluated by independent masked assessors, in a randomised order.

O que encontraram

Overall, 92 (41 %) participants fulfilled the TFOS DEWS II criteria for dry eye disease. The diagnostic performances for percentage meibomian gland dropout measurements (C-statistic range, 0.629 to 0.647) were significantly greater than qualitative meiboscale grading (C-statistic range, 0.547 to 0.560) for both instruments. The Youden-optimal diagnostic cut-off percentage gland dropout was > 20 % and optimal threshold meiboscale grade was > 1 for both superior and inferior eyelid measurements from the two instruments. Inter-instrument weighted Cohen's kappa coefficients for meiboscale grading were 0.734 and 0.682, respectively, and Bland-Altman biases (95 % limits of agreement) for percentage gland dropout were -1.8 % (-28.1 % to 24.4 %) and -1.5 % (-29.3 % to 26.2 %), respectively.

Por que isso importa

Despite exhibiting comparable discriminative abilities, meibography parameters obtained from the Keratograph and LipiView were not directly interchangeable. Overall, percentage gland dropout measurements demonstrated superior diagnostic performance, which would support their use in clinical trials and epidemiology studies, where practicable.

Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.

O estudo original

Diagnostic performance of qualitative and quantitative methods of meibomian gland dropout evaluation in dry eye disease: An investigator-masked, randomised crossover study

Michael T M Wang, Barry Power, Ally L Xue, Ji Soo Kim, Jennifer P Craig

Descritores MeSH

  • Diagnostic Techniques, Ophthalmological
  • Meibomian Glands
  • Dry Eye Syndromes
  • Male
  • Middle Aged
  • Prospective Studies
  • Tears
  • Reproducibility of Results
  • Single-Blind Method
  • Cross-Over Studies