Contrast of oropharyngeal leak pressure and clinical performance of I-gel™ and LMA ProSeal™ in patients: A meta-analysis
InfectologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
Conflicting outcomes have been reported for the i-gel™ and laryngeal mask airway (LMA) ProSeal™ in children and adults during general anesthesia. Randomized controlled trials (RCTs) that yielded wide contrast outcomes between i-gel™ and LMA ProSeal™ were included in this meta-analysis.
Two authors independently identified RCTs that compared i-gel™ with LMA ProSeal™ among patients receiving general anesthesia by performing searches in EMBASE, Cochrane, PubMed, and ScienceDirect. Discussion was adopted to resolve disagreements. Data were counted with Review Manger 5.3 and pooled by applying weighted mean difference (MD) and rlsk ratio (RR), and related 95% confidence intervals.
O que encontraram
A total of 33 RCTs with 2605 patients were included in the meta-analysis. I-gel™ provided a considerably lower oropharyngeal leak pressure [weighted average diversity (MD) = -1.53 (-2.89, -0.17), P = 0.03], incidence of blood staining on the supraglottic airway devices [RR = 0.44, (0.28, 0.69), P = 0.0003], sore throat [RR = 0.31 (0.18, 0.52), P<0.0001], and a short insertion time [MD = -5.61 (-7.71, -3.51), P<0.00001] than LMA ProSeal™. Compared with LMA ProSeal™, i-gel™ offered a significantly higher first-insertion success rate [RR = 1.03 (1.00, 1.06), P = 0.03] and ease of insertion [RR = 1.06 (1.01, 1.11), P = 0.03]. The gastric-tube-placement first insertion rate [RR = 1.04 (0.99, 1.10), P = 0.11], laryngospasm [RR = 0.76 (0.17, 3.31), P = 0.72], and cough [RR = 1.30 (0.49, 3.44), P = 0.60] between the two devices were similar.
Por que isso importa
Both devices could achieve a good seal to provide adequate ventilation. Compared with the used LMA ProSeal™, the i-gel™ was found to have fewer complications (blood stainning, sore throat) and offers certain advantages (short insertion time, higher first-insertion success rate and ease of insertion) in patients under general anesthesia.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Contrast of oropharyngeal leak pressure and clinical performance of I-gel™ and LMA ProSeal™ in patients: A meta-analysis
Yuan Tan, Jingyao Jiang, Rurong Wang
Descritores MeSH
- Pharyngitis
- Laryngeal Masks
- Child
- Adult
- Pain
- Humans
- Anesthesia, General
Veja também
Estudos relacionados
- Comparison of the Laryngeal Mask Airway ProSeal and the Streamlined Liner of the Pharynx Airway During General Anesthesia: A Systematic Review and Meta-analysis
- Supraglottic airway device versus tracheal tube for pediatric laparoscopic surgery-A systematic review and meta-analysis
- Airway Complications in Intubated Versus Laryngeal Mask Airway-Managed Dentistry: A Meta-Analysis
- Comparison of the efficacy of supraglottic airway devices in low-risk adult patients: a network meta-analysis and systematic review