Comparison of the effect of double-lumen endotracheal tubes and bronchial blockers on lung collapse in video-assisted thoracoscopic surgery: a systematic review and meta-analysis
PneumologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
This meta-analysis compared the quality of lung collapse and the resultant adverse reactions between the use of double-lumen endotracheal tubes (DLT) and bronchial blockers (BB) in minimally invasive thoracic surgery.
A search was performed in five bibliographic databases, namely PubMed, Springer, Medline, EMBASE, and Cochrane Library ignoring the original language, which identified five randomized controlled trials (RCTs) published on or before December 31, 2021. These studies were subsequently analyzed. All included studies compared the efficacy and safety of DLT and BB as a lung isolation technique in surgery. The methodological quality of each study was assessed by the Cochrane Collaboration's risk of bias tool. The quality of lung collapse and the malposition rate were adopted as the main outcome indicators. Alternatively, the intubation time and the incidence of postoperative sore throat were adopted as secondary indicators.
O que encontraram
When either DLT or BB were utilized in minimally invasive thoracic surgery, no differences were observed in the quality of lung collapse (odds ratio [OR], 1.00; 95% confidence interval [CI], 0.63 to 1.58), the intubation time (mean difference [MD], 0.06; 95% CI, -1.02 to 1.14), or the malposition rate (OR, 0.88; 95% CI, 0.37 to 2.06). However, the incidence of postoperative sore throat among patients treated with BB was significantly lower than that among patients treated with DLT (OR, 5.25; 95% CI, 2.55 to 10.75).
Por que isso importa
When utilized in minimally invasive thoracic surgery, the quality of lung collapse with DLT was identical to that with BB. However, patients treated with the latter demonstrated a significantly lower incidence of postoperative sore throat.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparison of the effect of double-lumen endotracheal tubes and bronchial blockers on lung collapse in video-assisted thoracoscopic surgery: a systematic review and meta-analysis
Ying-Ying Xiang, Qi Chen, Xi-Xi Tang, Lei Cao
Descritores MeSH
- Pulmonary Atelectasis
- Pharyngitis
- One-Lung Ventilation
- Humans
- Intubation, Intratracheal
- Thoracic Surgery, Video-Assisted
Veja também
Estudos relacionados
- Facilitating Lung Collapse for Thoracoscopic Surgery Utilizing Endobronchial Airway Occlusion Preceded by Pleurotomy and One-minute Suspension of Two-lung Ventilation
- Non-ventilated lung deflation during one-lung ventilation with a double-lumen endotracheal tube: a randomized-controlled trial of occluding the non-ventilated endobronchial lumen before pleural opening
- Double-lumen endotracheal tubes and bronchial blockers exhibit similar lung collapse physiology during lung isolation
- Left-Sided Double-Lumen Tube vs EZ-Blocker for One-Lung Ventilation in Thoracic Surgery: A Systematic Review and Meta-Analysis