Comparison of the coverage and rotation of asymmetrical and symmetrical tibial components: a systematic review and meta-analysis
CirurgiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
An optimized fit of the tibial component to the resection platform and correct rotational alignment are critical for successful total knee arthroplasty (TKA). However, there remains controversy regarding the superiority of symmetric tibial component versus asymmetric tibial component. The objective of this systematic review and meta-analysis was to evaluate the current evidence for comparing the coverage and rotation of asymmetrical and symmetrical tibial component.
We searched potentially relevant studies form PubMed, Web of science, Embase, Cochrane Central Register of Controlled Trials (CENTRAL), and China National Knowledge Infrastructure (CNKI), up to 1 March 2023. Data extraction and quality assessment were performed by two independent reviewers. Meta-analysis was conducted using Review Manager 5.4.
O que encontraram
Sixteen articles were identified. Compared to symmetric tibial component, asymmetric tibial component increased the coverage of the proximal tibial cut surface (MD, -2.87; 95%CI, -3.45 to -2.28; P < 0.00001), improved the prevalence of tibial baseplate underhang (OR, 0.16; 95%CI, 0.07 to 0.33; P < 0.00001) and malrotation (OR, 0.13; 95%CI, 0.02 to 0.90; P = 0.04), and reduced the degree of tibial component rotation (MD, -3.11; 95%CI, -5.76 to -0.47; P = 0.02). But there was no statistical significance for improving tibial baseplate overhang (OR, 0.58; 95%CI, 0.08 to 3.97; P = 0.58). Additionally, no revision had occurred for the two tibial components in the included studies.
Por que isso importa
The current evidence shows asymmetric tibial component offer advantages in terms of coverage and rotation compared with symmetric tibial component in TKA.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparison of the coverage and rotation of asymmetrical and symmetrical tibial components: a systematic review and meta-analysis
Ziming Zhang, Tiantian Zhang, Le Zhang, Zehua Chen, Haoming Zhao, Jianjun Kuang, Liang Ou
Descritores MeSH
- Knee Joint
- Knee Prosthesis
- Arthroplasty, Replacement, Knee
- Tibia
- Treatment Outcome
- Humans
- Osteoarthritis, Knee
- Prosthesis Design
- Rotation
- Range of Motion, Articular
Veja também
Estudos relacionados
- Knee kinematics are not different between asymmetrical and symmetrical tibial baseplates in total knee arthroplasty: A fluoroscopic analysis of step-up and lunge motions
- Influence of component design on in vivo tibiofemoral contact patterns during kneeling after total knee arthroplasty: a systematic review and meta-analysis
- Measured resection, gap balancing, and force-sensor-guided total knee arthroplasty result in different femoral and tibial rotational alignment but similar combined knee rotation and clinical outcomes: a randomized controlled trial
- Does a medial congruent bearing in total knee arthroplasty compromise fixation? A randomized controlled trial