Comparison of procedural efficacy, balloon nadir temperature, and incidence of phrenic nerve palsy between two cryoballoon technologies for pulmonary vein isolation: A systematic review and meta-analysis
CardiologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
In May 2020, a novel cryoballoon system (POLARx; Boston Scientific) became available for catheter ablation of atrial fibrillation (AF). The design of the cryoballoon is comparable to the Arctic Front Advance Pro (AFA-Pro; Medtronic), but it is more compliant during freezing. We compared the procedural efficacy, biophysical parameters, and risk of phrenic nerve palsy (PNP) between the two cryoballoons.
Embase, MEDLINE, Web of Science, Cochrane, and Google Scholar databases were searched until June 1, 2021 for relevant studies comparing POLARx versus AFA-Pro in patients undergoing pulmonary vein isolation (PVI) for AF.
O que encontraram
A total of four studies, involving 310 patients were included. There was no difference between the two groups for outcomes regarding procedural efficacy: acute PVI (odds ratio [OR]: 0.43; 95% confidence interval [CI]: 0.06 to 3.03; p = .40), procedure time (mean difference [MD]: 8.15 min; 95% CI: -8.09 to 24.39; p = .33), fluoroscopy time (MD: 1.32 min; 95% CI: -1.61 to 4.25; p = .38) and ablation time (MD: 1.00 min; 95% CI: -0.20 to 2.20; p = .10). The balloon nadir temperature was lower for all individual pulmonary veins (PV) in POLARx compared with AFA-Pro (MD: -9.74°C, -9.98°C, -6.72°C, -7.76°C, for left superior PV, left inferior PV, right superior PV, and right inferior PV, respectively; all p < .001). The incidence of PNP was similar between groups (OR: 0.79; 95% CI: 0.22 to 2.85; p = .72).
Por que isso importa
In AF patients undergoing PVI, POLARx and AFA-Pro had a similar procedural efficacy. Balloon nadir temperatures were lower with POLARx, however, the incidence of PNP was similar.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparison of procedural efficacy, balloon nadir temperature, and incidence of phrenic nerve palsy between two cryoballoon technologies for pulmonary vein isolation: A systematic review and meta-analysis
Amira Assaf, Rohit Bhagwandien, Tamas Szili-Torok, Sing-Chien Yap
Descritores MeSH
- Catheter Ablation
- Cryosurgery
- Pulmonary Veins
- Atrial Fibrillation
- Phrenic Nerve
- Treatment Outcome
- Temperature
- Incidence
- Humans
- Paralysis
Veja também
Estudos relacionados
- Comparison for the efficacy and safety of time-to-isolation protocol and conventional protocol of cryoballoon in the treatment of atrial fibrillation: a meta-analysis and systematic review
- Efficacy and safety of a novel size-adjustable cryoballoon vs conventional cryoballoon: Results from the CONTRAST-CRYO II randomized clinical trial
- Cryoballoon Pulmonary Vein Isolation With Versus Without Additional Right Atrial Linear Ablation for Persistent Atrial Fibrillation: The CRALAL Randomized Clinical Trial
- Randomized prospective evaluation of same-day discharge after cryoballoon ablation of atrial fibrillation: results of the EASY PVI study