Comparison of a portable, pneumotach flow-sensor-based spirometer (Spirofy™) with the vitalograph alpha Touch™ spirometer in evaluating lung function in healthy individuals, asthmatics, and COPD patients-a randomized, crossover study
PneumologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
Spirofy™ is India's first portable, pneumotach flow-sensor-based digital spirometer developed to diagnose asthma and chronic obstructive pulmonary disease (COPD). In this study, we compared the performance of the Spirofy™ device with that of the Vitalograph Alpha Touch™ spirometer in measuring the lung capacities of healthy individuals, asthmatics, and COPD patients. We also assessed the inter-device variability between two Spirofy™ devices.
In a randomized, three-way crossover, open-label study, we measured the differences in forced expiratory volume in the first second (FEV1) and forced vital capacity (FVC) between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. A proportion of the FEV1/FVC ratio distribution of < 0.7 was used to compare the diagnostic accuracies of the Spirofy™ with Vitalograph™ Alpha Touch™ spirometers.
CTRI/2021/09/036492 (Clinical Trials Registry - India).
O que encontraram
Ninety subjects participated in this study. The mean ± SD FVC values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 2.60 ± 1.05 L, 2.64 ± 1.04 L, and 2.67 ± 1.04 L, respectively. The mean ± SD FEV1 values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 1.87 ± 0.92 (L), 1.88 ± 0.92 (L), and 1.93 ± 0.93 (L), respectively. A significant positive correlation was found between the FVC and FEV1 values recorded by Vitalograph Alpha Touch™, Spirofy™ 1, and Spirofy™ 2. As compared to Vitalograph Alpha Touch™, the Spirofy™ device showed good sensitivity (97%), specificity (90%), and overall accuracy (93.3%) at an FEV1/FVC ratio < 0.7. No inter-device variability was observed between the two Spirofy™ devices.
Por que isso importa
Spirofy™ is a portable and easy-to-use device and is as accurate as the standard Vitalograph Alpha Touch™ spirometer for the diagnosis of COPD and asthma.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparison of a portable, pneumotach flow-sensor-based spirometer (Spirofy™) with the vitalograph alpha Touch™ spirometer in evaluating lung function in healthy individuals, asthmatics, and COPD patients-a randomized, crossover study
Deepak Talwar, S Balamurugan, Mahavir Modi, Sundeep Salvi, Meena Lopez, Rushika Shah, Abhijit Vaidya, Monica Barne, Sapna Madas, Nandan Kulkarni et al.
Descritores MeSH
- Asthma
- Cross-Over Studies
- Pulmonary Disease, Chronic Obstructive
- Spirometry
- Humans
- India
- Male
- Middle Aged
- Vital Capacity
- Adult
Veja também
Estudos relacionados
- Reliability and usability of a portable spirometer compared to a laboratory spirometer
- Early detection of chronic obstructive pulmonary disease in primary care: a randomised controlled trial
- Bronchodilators for hyperinflation in COPD associated with biomass smoke: clinical trial
- Conectando muco em várias doenças respiratórias crônicas: revisão sistemática e meta-análise