
Comparison between compressed sensing and segmented cine cardiac magnetic resonance: a meta-analysis
Radiologia e ImagemPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
Highly accelerated compressed sensing cine has allowed for quantification of ventricular function in a single breath hold. However, compared to segmented breath hold techniques, there may be underestimation or overestimation of LV volumes. Furthermore, a heterogeneous sample of techniques have been used in volunteers and patients for pre-clinical and clinical use. This can complicate individual comparisons where small, but statistically significant differences exist in left ventricular morphological and/or functional parameters. This meta-analysis aims to provide a comparison of conventional cine versus compressed sensing based reconstruction techniques in patients and volunteers.
Two investigators performed systematic searches for eligible studies using PubMed/MEDLINE and Web of Science to identify studies published 1/1/2010-3/1/2021. Ultimately, 15 studies were included for comparison between compressed sensing cine and conventional imaging.
O que encontraram
Compared to conventional cine, there were small, statistically significant overestimation of LV mass, underestimation of stroke volume and LV end diastolic volume (mean difference 2.65 g [CL 0.57-4.73], 2.52 mL [CL 0.73-4.31], and 2.39 mL [CL 0.07-4.70], respectively). Attenuated differences persisted across studies using prospective gating (underestimated stroke volume) and non-prospective gating (underestimation of stroke volume, overestimation of mass). There were no significant differences in LV volumes or LV mass with high or low acceleration subgroups in reference to conventional cine except slight underestimation of ejection fraction among high acceleration studies. Reduction in breath hold acquisition time ranged from 33 to 64%, while reduction in total scan duration ranged from 43 to 97%.
Por que isso importa
LV volume and mass assessment using compressed sensing CMR is accurate compared to conventional parallel imaging cine.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparison between compressed sensing and segmented cine cardiac magnetic resonance: a meta-analysis
Jason Craft, Yulee Li, Niloofar Fouladi Nashta, Jonathan Weber
Descritores MeSH
- Heart
- Magnetic Resonance Imaging
- Heart Ventricles
- Humans
- Magnetic Resonance Spectroscopy
- Breath Holding
Veja também
Estudos relacionados
- Guidelines for the Evaluation of Prosthetic Valve Function With Cardiovascular Imaging: A Report From the American Society of Echocardiography Developed in Collaboration With the Society for Cardiovascular Magnetic Resonance and the Society of Cardiovascular Computed Tomography
- Cardiovascular magnetic resonance native T1 mapping in Anderson-Fabry disease: a systematic review and meta-analysis
- Four-dimensional flow cardiovascular magnetic resonance in tetralogy of Fallot: a systematic review
- Impacto de uma intervenção de treinamento intervalado de alta intensidade de 12 semanas na estrutura e função cardíaca após COVID-19 no seguimento de 12 meses