Comparing Equil patch versus traditional catheter insulin pump in type 2 diabetes using continuous glucose monitoring metrics and profiles
EndocrinologiaResumo original em inglês
O que aconteceu
It is not clear whether there are differences in glycemic control between the Equil patch and the MMT-712 insulin pump. Our objective was to compare two types of insulin pumps in the treatment of type 2 diabetes mellitus (T2DM), using continuous glucose monitoring (CGM) metrics and profiles.
This was a randomized case-crossover clinical trial. Participants were hospitalized and randomly allocated to two groups and underwent two types of insulin pump treatments (group A: Equil patch-Medtronic MMT-712 insulin pump; group B: Medtronic MMT-712-Equil patch insulin pump) separated by a 1-day washout period. Glycemic control was achieved after 7-8 days of insulin pump therapy. Each patient received CGM for 5 consecutive days (from day 1 to day 5). On day 3 of CGM performance, the Equil patch insulin pump treatment was switched to Medtronic MMT-712 insulin pump treatment at the same basal and bolus insulin doses or vice versa. CGM metrics and profiles including glycemic variability (GV), time in range (TIR, 3.9-10.0 mmol/L), time below range (TBR, 10.0 mmol/L), and postprandial glucose excursions, as well as incidence of hypoglycemia.
O que encontraram
Forty-six T2DM patients completed the study. There was no significant difference in parameters of daily GV and postprandial glucose excursions between the Equil patch insulin pump treatment and the Medtronic insulin pump treatment. Similarly, there was no between-treatment difference in TIR, TBR, and TAR, as well as the incidence of hypoglycemia.
Por que isso importa
The Equil patch insulin pump was similar to the traditional MMT-712 insulin pump in terms of glycemic control. Equil patch insulin pump is a reliable tool for glycemic management of diabetes mellitus.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparing Equil patch versus traditional catheter insulin pump in type 2 diabetes using continuous glucose monitoring metrics and profiles
Yu-Jiao Li, Zi-Yue Shao, Yun-Qing Zhu, Da-Shuang Chen, Jian Zhu
Descritores MeSH
- Hypoglycemia
- Diabetes Mellitus, Type 2
- Glucose
- Humans
- Hypoglycemic Agents
- Insulin
- Blood Glucose Self-Monitoring
- Cross-Over Studies
- Continuous Glucose Monitoring
- Catheters
Veja também
Estudos relacionados
- Efficacy and safety of continuous glucose monitoring on glycaemic control in patients with chronic pancreatitis and insulin-treated diabetes: A randomised, open-label, crossover trial
- Extended use of real-time continuous glucose monitoring in adults with insulin-requiring type 2 diabetes: Results from the first 26 weeks of the 2GO-CGM trial
- Does fully closed-loop automated insulin delivery improve glycaemic control in patients with type 2 diabetes? A meta-analysis of randomized controlled trials
- Rapid improvements in glycemic management with use of continuous glucose monitoring in adults with type 2 diabetes treated with basal insulin: 3-month analysis of the MOBILE study