Comparative 30-day echocardiographic outcomes of Myval vs. Sapien and Evolut THVs: insights from LANDMARK trial
CardiologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
Several factors, including device design, annulus size, and sizing strategies, influence transcatheter heart valve (THV) haemodynamic outcomes in patients with aortic stenosis (AS). This sub-study evaluates early (30-day) echocardiographic outcomes of the Myval, Sapien, and Evolut THV series, focusing on haemodynamic performance and valve durability.
ClinicalTrials.gov: NCT04275726, EudraCT number 2020-000,137-40.
O que encontraram
The LANDMARK trial is a prospective, randomised, multicentre, open-label, non-inferiority trial comparing 384 patients implanted with Myval THV series to 384 receiving Sapien and Evolut THV series. Haemodynamic assessments followed Valve Academic Research Consortium-3 recommendations. At 30-day, haemodynamic device success rates were 85.9%, 77.8, and 85.4% for Myval, Sapien, and Evolut THV series, respectively (PMyval-Sapien = 0.03 and PMyval-Evolut = 0.98). Significant improvements in peak aortic flow velocity, pressure gradients, effective orifice area (EOA), Doppler velocity index (DVI), and cardiac indices were observed across all groups, except for unchanged left ventricular ejection fraction. Moderate prosthesis-patient mismatch (PPM) was less frequent with Myval THV series(11.3%) vs. Sapien THV series(21.8%), but higher than Evolut THV series (5.3%) (PMyval-Sapien = 0.0024, PMyval-Evolut = 0.0396), while severe PPM showed no significant differences (4.2% vs. 6.3% vs. 1.8%; PMyval-Sapien = 0.394, PMyval-Evolut = 0.2438). Rates of ≥ moderate paravalvular leak (PVL) were lower in Myval (3.5%), and Sapien (1.7%) compared with Evolut THV series (8.3%) (PMyval-Sapien = 0.3769, PMyval-Evolut = 0.0336). Myval THV series required minimal oversizing compared with Evolut THV series (P < 0.0001).
Por que isso importa
The Myval THV series demonstrates short-term haemodynamic performance comparable to Evolut THV series and superior to Sapien THV series. Including intermediate sizes minimizes oversizing, underscoring its potential as an alternative for TAVI patients. Long-term follow-up is necessary to confirm these findings.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Comparative 30-day echocardiographic outcomes of Myval vs. Sapien and Evolut THVs: insights from LANDMARK trial
Osama Soliman, Elfatih A Hasabo, Niels van Royen, Ignacio Jesus Amat-Santos, Martin Hudec, Matjaz Bunc, Alexander IJsselmuiden, Peep Laanmets, Daniel Unic, Béla Merkely et al.
Descritores MeSH
- Aortic Valve Stenosis
- Heart Valve Prosthesis
- Echocardiography, Doppler
- Echocardiography
- Transcatheter Aortic Valve Replacement
- Male
- Prospective Studies
- Prosthesis Design
- Time Factors
- Treatment Outcome
Veja também
Estudos relacionados
- Balloon- Versus Self-Expanding Transcatheter Valves for Failed Small Surgical Aortic Bioprostheses: 3-Year Results of the LYTEN Trial
- Revisão sistemática atualizada e metanálise comparando as valvas cardíacas transcateter Myval versus Sapien 3 em estenose aórtica grave: foco na mortalidade precoce e em um ano
- Five-Year Clinical Outcomes and Durability of a Self-Expanding Transcatheter Heart Valve With Intra-Annular Leaflets
- 5 anos de seguimento da substituição da válvula aórtica transcateter com válvulas intra-anular auto-expansíveis disponíveis comercialmente versus