Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses
OrtopediaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
Short-stemmed endoprostheses were developed to implement proximal load transmission and thus avoid stress-shielding in the proximal femur. Various prosthesis systems have been developed, which are discussed in the literature regarding stress shielding, clinical outcome, and long-term implant stability.
In this prospective randomized study, 52 patients (27 male, 25 female; average age 60.8 years) with conservatively unsuccessfully treated coxarthrosis were implanted with either a Nanos™ or Optimys™ short-stem prosthesis. Assessment included Gruen-zone based DEXA examinations immediately postoperatively and at one year to evaluate bone mineral density (BMD) and stress shielding, along with clinical outcomes using the Harris Hip Score (HHS). Radiographic measurements included offset (OFF), caput-collum-diaphyseal angle (CCD), leg length (LL), stem migration and inclination, and the occurrence of radiolucent lines (RL), assessed preoperatively, postoperatively, and at 12 months.
Ib.
O que encontraram
DEXA showed differing stress-shielding profiles between stem types, favoring Optimys™ for BMD preservation. The Nanos™ group exhibited significantly greater BMD reduction in Gruen zones 1 (- 10.1%; p = 0.001), 4 (- 3.2%; p = 0.02), and 7 (- 21.3%; p = 0.001), whereas Optimys™ showed a significant decrease only in zone 7 (- 16.2%; p = 0.001). Although OFF, CCD, and LL changed significantly within groups postoperatively (p 0.05). Stem migration remained clinically irrelevant in both groups. A statistically significant intra-group change was observed only in the Optimys™ group (Nanos™: 1.7 mm, p = 0.13; Optimys™: 2.5 mm, p = 0.01). Similarly, a small but statistically significant change in stem inclination was observed within both groups (Nanos™: 2.2°, p = 0.002; Optimys™: 1.5°, p = 0.01). Clinical improvement as measured by the Harris Hip Score (HHS) was excellent in both groups, with no significant differences between systems (Nanos™ pre/post: 52.0 / 98.0; Optimys™ pre/post: 51.6 / 97.0; both p < 0.001).
Por que isso importa
When compared, the Optimys stem demonstrated reduced stress shielding through improved proximal load transmission, resulting in significantly better preservation of bone mineral density in the proximal femur.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
Clinically relevant differences in stress shielding between two short-stemmed femoral prostheses
Felix Werneburg, Annabell Herntrich, Julia Dietz, David Wohlrab, Natalia Gutteck, Delank Karl-Stefan, Alexander Zeh
Descritores MeSH
- Osteoarthritis, Hip
- Prosthesis Design
- Hip Prosthesis
- Arthroplasty, Replacement, Hip
- Male
- Middle Aged
- Aged
- Stress, Mechanical
- Absorptiometry, Photon
- Bone Density
Veja também
Estudos relacionados
- Preservação de osso periprotético com haste curta fixa de collum versus hastes fixas metafisárias convencionais na artroplastia total da anca: um ensaio controlado randomizado com seguimento de 2 anos em 61 pacientes
- Similar performance with electrochemically deposited hydroxyapatite coating versus plasma porous spray in press-fit acetabular components for primary total hip arthroplasty: a randomised controlled study of early outcomes with DEXA
- Long-term evaluation of periprosthetic bone changes in ultra-short versus conventional stems in total hip arthroplasty: a 10-year follow-up of a randomised controlled trial
- Similar femoral stem fixation but less metaphyseal loss of bone mineral density with a taper-wedge design and diaphyseal bone preservation with a long and round-tapered design: a 5-year randomized RSA and DXA study of 50 patients