A fully artificial pancreas versus a hybrid artificial pancreas for type 1 diabetes: a single-centre, open-label, randomised controlled, crossover, non-inferiority trial
EndocrinologiaPor ScienceMed · · a partir do resumo dos autores
Resumo original em inglês
O que aconteceu
For people with type 1 diabetes, there is currently no automated insulin delivery system that does not require meal input. We aimed to assess the efficacy of a novel faster-acting insulin aspart (Fiasp) plus pramlintide fully closed-loop system that does not require meal input.
In this open-label, randomised controlled, crossover, non-inferiority trial we compared the Fiasp (Novo Nordisk, Bagsværd, Denmark) plus pramlintide closed-loop system with no meal input (fully artificial pancreas) and the Fiasp-alone closed-loop system with precise carbohydrate counting (hybrid artificial pancreas). Adults (≥18 years) who had a clinical diagnosis of type 1 diabetes for at least 12 months, had glycated haemoglobin 12% or lower, and had been on insulin pump therapy for at least 6 months were enrolled at McGill University Health Centre, Montreal, QC, Canada. The Fiasp plus pramlintide fully closed-loop system delivered pramlintide in a basal-bolus manner with a fixed ratio of 10 μg:U relative to insulin. A research staff member counted the carbohydrate content of meals to input in the hybrid closed-loop system. Participants completed the two full-day crossover interventions in a random order allocated by a computer-generated code implementing a blocked randomisation (block size of four). The primary outcome was the percentage of time spent within the glucose target range (3·9-10·0 mmol/L), with a 6% non-inferiority margin, assessed in all participants who completed both interventions. This trial is registered with ClinicalTrials.gov, NCT03800875.
Diabetes Canada.
O que encontraram
Between Feb 8, 2019, and Sept 19, 2020, we enrolled 28 adults, of whom 24 completed both interventions and were included in analyses. The percentage of time spent in the target range was 74·3% (IQR 61·5-82·8) with the fully closed-loop system versus 78·1% (66·3-87·5) with the hybrid Fiasp-alone closed-loop system (paired difference 2·6%, 95% CI -2·4 to 12·2; non-inferiority p=0·28). Eight (33%) participants had at least one hypoglycaemia event (<3·3 mmol/L) with the fully closed-loop system compared with 14 (58%) participants with the hybrid closed-loop system (2200-2200 h). Non-mild nausea was reported by three (13%) participants and non-mild bloating by one (4%) participant with the fully closed-loop system compared with zero participants with the hybrid closed-loop system.
Por que isso importa
The Fiasp plus pramlintide fully closed-loop system was not non-inferior to the Fiasp-alone hybrid closed-loop system for the overall percentage of time in the glucose target range. However, participants still spent a high percentage of time within the target range with the fully-closed loop system. Outpatient studies comparing the fully closed-loop hybrid systems with patient-estimated, rather than precise, carbohydrate counting are warranted.
Resumo publicado sob licença CC BY. A tradução para o português ainda não chegou a este estudo.
O estudo original
A fully artificial pancreas versus a hybrid artificial pancreas for type 1 diabetes: a single-centre, open-label, randomised controlled, crossover, non-inferiority trial
Michael A Tsoukas, Dorsa Majdpour, Jean-François Yale, Anas El Fathi, Natasha Garfield, Joanna Rutkowski, Jennifer Rene, Laurent Legault, Ahmad Haidar
Descritores MeSH
- Islet Amyloid Polypeptide
- Diabetes Mellitus, Type 1
- Hypoglycemic Agents
- Insulin
- Insulin Infusion Systems
- Pancreas, Artificial
- Insulin, Long-Acting
- Humans
- Hypoglycemia
- Cross-Over Studies
Veja também
Estudos relacionados
- Efficacy and Safety of Tubeless Hybrid Closed-Loop Systems in Type 1 Diabetes: A Meta-Analysis of Randomized Controlled Trials
- Hybrid closed-loop glucose control compared with sensor augmented pump therapy in older adults with type 1 diabetes: an open-label multicentre, multinational, randomised, crossover study
- Efficacy and safety of automated insulin delivery system in very young children with type 1 diabetes: A systematic review and meta-analysis of randomized controlled trials
- Efficacy and safety of once-weekly insulin versus once-daily insulin in patients with type 1 and type 2 diabetes mellitus: an updated meta-analysis of randomized controlled trials
- Comparing Automated Lyumjev® Delivery with Carbohydrate Counting, Qualitative Meal-Size Estimation, and Fully Closed-Loop in Type 1 Diabetes: A Pilot, Randomized, Crossover Trial